Model
|
19501-K
|
AC Output Voltage
|
Range
|
0.10 kV~10.00 kV, steps 0.01 kV
|
Voltage Accuracy
|
±(1% of setting + 0.5% full Scale)
|
Load Regulation
|
±(1% of setting + 0.5% full scale)
|
Frequency
|
50Hz, 60Hz ± 0.1%, sine wave
|
Measurement
|
V-display Accuracy
|
± (1% of reading + 0.5% full scale), 10V resolution
|
Cutoff Current
|
0.01µA~300.0µA
|
Leakage Current Meter
|
30µA : 0.50µA~29.99µA ; 300µA : 30.00~300.0µA
|
|
Accuracy ± (1% of reading + 2% full scale)
|
Partial Discharge Detector
|
Range
|
200pC: 1.0 pC~200 pC, 0.1 pC resolution ; 2000pC: 10 pC~2000 pC, 1 pC resolution
|
Accuracy
|
± (1% of reading + 0.5% full scale)
|
Test TimerTest time
|
0.3~99.9 sec., steps 0.1 sec.
|
Ramp/Fall time
|
0.1~9.9 sec., steps 0.1 sec
|
PD detection delay time:
|
0~9.9 sec., steps 0.1 sec.
|
HVCC
|
* Added HV & RTN contact terminal, test current <10mA and open voltage is 5Vdc typ.
|
* Check function to select ON or OFF
|
|
36 pins connector, all input/output are negative true logic and optically isolated
|
|
open collector signals. (General-speed photo-coupler used)
|
Handler interface
|
* all outputs must be pulled up with 10kohm resistor to +VEXT (external power supply)
|
|
* all input optic-diode must be series with current limit (10mA±4mA for +3V ~ +26V) circuit
|
Remote Interface
|
RS-232
|
The programming language is SCPI
|
USB (B-type)
|
Meet USB TMC
|
LAN
|
Support 10M/100M Ethernet
|
Memory Storage
|
200 instrument setups
|
|
Test parameters, result and waveform (BMP) storage. (EXP. function)
|
|
One memory of test procedure and parameter can be storage/recall Backup/restore all memory data to USB flash
|
USB Flash Drive (A-type)
|
|
|
Support USB Flash up to 32GB in size
|
General
|
Specifications Range
|
18 to 28 ℃ (64 to 82℉), 70% RH
|
Operable Range
|
0 ℃ to 45 ℃, 15% to 95% RH @ ≦40 ℃ and no condensation
|
Storage Range
|
-10℃ to 50 ℃, ≦ 80% RH
|
Power Requirement
|
100 Vac ~ 240 Vac, 50/60 Hz
|
Power Consumption
|
No Load : <150W, Rated Load : <400W
|
Dimension (WxHxD)
|
Host: 428x176x500 mm / 16.9x6.9x19.7 inches
|
HV Box: 203x307x200 mm / 8x12.1x7.9 inches
|
Weight
|
Host: 20.5 kg / 45.19 lbs
|
HV Box: 13.2 kg / 29.10 lbs
|
|
|
Model
|
A195001
|
Output Voltage
|
1~100V
|
Polarity
|
Negative
|
Range
|
1,5,100pC in 200pC range ; 20,100,2000pC in 2000pC range
|
Accuracy
|
± (3% of reading + 0.5pC)
|
Rise Time
|
<50nS
|
Pulse Repetition
|
100Hz
|
Injection Capacitance
|
1pF
|
Operable Range
|
0℃ to 45℃, 15% to 95% RH@ ≦40℃ and no condensation
|
Storage Range
|
-10 to 50℃, ≦ 80% RH
|
Power Supply
|
9V battery
|
Current Consumption
|
50mA max.
|
Dimension (WxHxD)
|
100 x 58 x 151.7 mm / 3.94 x 2.28 x 5.97 inch
|
Weight
|
Approx. 1.25kg
|